The Agilent/Keysight 54200D is a 50 MHz digitizing oscilloscope built for signal visualization and precision measurement across electronics testing, research, and development environments. Dual 200 MSa/s digitizers capture transient and repetitive signals with 8-bit vertical resolution across two channels. The instrument excels in digital logic applications through advanced state, missing bit, and extra bit triggering—each with parallel and serial logic qualification. Pre-trigger display captures signal activity before the trigger event, while multiple acquisition modes (infinite persistence, envelope, average) support both live and stored waveform analysis.
Technical Specifications
Vertical System
• Bandwidth: 50 MHz single-shot (post-capture interpolation)
• Full-scale range: 40 mV to 40 V continuously calibrated, 2-digit resolution
• Gain accuracy: ±2% of full-scale
• Vertical resolution: 8-bit A/D conversion
• Input coupling: DC, AC, GND
• Maximum safe input: 140 V peak
Horizontal System
• Time base: 50 ns to 10 s per full scale (10 divisions), 1-2-5 sequence
Acquisition
• Sampling rate: 200 MSa/s per channel
• Memory depth: 1 kbyte per channel
• Channels: 2
• Pre-trigger storage: 1 screen diameter
Trigger System
• State trigger clock: 25 MHz max (single phase), 20 MHz max (multiplexed)
• Master-slave timing: Master must follow slave by ≥10 ns, precede next slave by ≥50 ns
• TTL/ECL preset capabilities for automatic vertical range, offset, and trigger level configuration
– Key Features
• State triggering up to 27 channels with parallel and serial logic qualification
• Missing bit and extra bit trigger modes for protocol-specific applications
• Automatic waveform measurements on pulse parameters
• Time and voltage cursors for live and stored waveform measurements
• Auto-scaling for rapid signal characterization
• Pre-trigger delay: 0 seconds (time data present after clock transition)
– Typical Applications
• Digital logic signal analysis and validation
• Protocol development and debugging
• Multi-channel timing analysis in synchronous systems
• Transient capture with protocol-aware triggering
– Compatibility & Integration
• HP-IB (IEEE-488) interface for remote control and programming of all front-panel functions
• Power input: 115 VAC, −22% to +10% (90–127 VAC)