The Agilent/Keysight 85102C is a test set component engineered for precision S-parameter measurements within the 8510 series network analyzer platform. Operating across 45 MHz to 110 GHz when configured with compatible sources and accessories, it delivers two-port network characterization with magnitude, phase, and group delay analysis. The modular design enables flexible system configuration tailored to specific measurement requirements.
Technical Specifications
• Frequency Range: 45 MHz to 110 GHz (system-dependent on source and test set configuration)
• S-Parameter Measurements: Two-port networks with magnitude, phase, and group delay; optional time-domain analysis via inverse Fourier transform
• Calibration: Full two-port calibration with sliding loads; compatible with 85052B 3.5 mm calibration kits
• Receiver Dynamic Range: Up to 106 dB
• System Dynamic Range: Up to 93 dB (frequency and configuration dependent)
• Measurement Resolution:
– Magnitude: 0.001 dB
– Phase: 0.01 degrees
– Group Delay: 0.01 nanoseconds
• Reference Power (Port 1): Typically +2 dBm or −2 dBm (frequency-dependent)
• Port 2 Power Range: −66 dBm to −95 dBm (minimum); +20 dBm to −4 dBm (maximum, frequency-dependent)
• IF Signal: 20 MHz
– Key Features
• J10 TEST SET INTERCONNECT port for system integration
• Standard RF input/output ports for device under test connection
• GPIB control interface
• 20 MHz output on rear panel for diagnostic and calibration functions
– Typical Applications
• Microwave component characterization
• Two-port network analysis and validation
• Integrated system S-parameter measurement
– Compatibility & Integration
Designed for integration with the Agilent/Keysight 8510C Vector Network Analyzer. Compatible with various RF sources and system accessories to achieve extended frequency coverage and measurement capabilities.