The Agilent/Keysight 86142A is a benchtop optical spectrum analyzer engineered for high-resolution spectral analysis across 600 nm to 1700 nm. It delivers exceptional wavelength accuracy—±0.025 nm (1510–1570 nm) and ±0.035 nm (1570–1640 nm) after internal calibration—and maintains ±0.003 nm tuning repeatability. The instrument achieves –90 dBm sensitivity in the 1250–1610 nm band and handles maximum input power of +30 dBm total across the full spectrum. Eight selectable resolution bandwidths range from 0.06 nm to 10 nm FWHM, enabling precise characterization of DWDM systems, EDFA performance, and passive optical components. Dynamic range reaches –70 dB in the 1550 nm band at ±0.5 nm span with 0.1 nm resolution. Scale fidelity holds ±0.05 dB characteristic for fixed reference level measurements. Wavelength reproducibility of ±0.002 nm and polarization dependence as low as ±0.05 dB support demanding R&D and production test environments demanding accuracy and stability.
Technical Specifications
• Wavelength Range: 600 nm to 1700 nm
• Wavelength Accuracy: ±0.025 nm (1510–1570 nm), ±0.035 nm (1570–1640 nm) after internal calibration; ±0.5 nm absolute (factory, 2-year cycle); ±10 pm with wavelength calibrator option
• Wavelength Reproducibility: ±0.002 nm
• Tuning Repeatability: ±0.003 nm
• Span Range: 0.2 nm to full range, continuously variable with zero span capability
• Resolution Bandwidth (FWHM): 0.06 nm, 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm, 10 nm
• Sensitivity: –60 dBm (600–750 nm); –75 dBm (750–1250 nm); –90 dBm (1250–1610 nm); –80 dBm (1610–1700 nm)
• Maximum Input Power: +30 dBm total; +15 dBm per channel (600–1000 nm); +12 dBm per channel (1000–1525 nm)
• Dynamic Range: –70 dB at 1550 nm (0.1 nm resolution, ±0.5 nm span, chop mode on); –55 dB characteristic at ±0.4 nm (±50 GHz)
• Scale Fidelity: ±0.05 dB characteristic (fixed reference level); ±0.07 dB characteristic (multiple reference levels)
• Absolute Amplitude Accuracy: ±0.35 dB at 1550 nm, –20 dBm, 20–30 °C
• Polarization Dependence: ±0.05 dB to ±0.25 dB (resolution ≥0.2 nm, 9/125 µm fiber)
– Key Features
• Eight independent resolution bandwidth settings for flexible trade-off between spectral detail and measurement speed
• Sub-picometer wavelength repeatability enables precise channel tracking in DWDM systems
• High sensitivity across the entire measurement band minimizes averaging requirements
• Safe power handling to +30 dBm within any 10 nm spectral segment
• Calibration options including user-defined references and external wavelength calibrator for sub-10 pm accuracy
– Typical Applications
• Dense Wavelength Division Multiplexing (DWDM) system and component testing
• Erbium-doped fiber amplifier (EDFA) performance characterization
• Passive optical component analysis
• Optical research and development
• Production-line optical measurement