The Fluke 1362S is a high-precision 6.5 digit VXI multimeter module designed for automated test systems. It offers accurate voltage, current, and resistance measurements. Ideal for demanding applications requiring high speed and accuracy in a modular instrument format.

Request CalibrationRequest Repair
Product Starting From:

$1,650.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 1362S
Equipment info

The Fluke/Wavetek/Datron 1362S is a 6.5-digit VXI multimeter module engineered for precision measurement within automated test systems. This C-size, single-slot VXIbus card delivers selectable resolution from 4.5 to 6.5 digits across DC voltage, AC voltage, resistance, and current measurements. Read rates reach 1000 readings per second at 4.5-digit resolution, stepping down to 5 readings/sec at 6.5-digit resolution for DC voltage. The module stores measurement blocks of 1 to 1000 readings and supports both front-panel and VXI backplane triggering via synchronous and asynchronous protocols. Programming is handled through SCPI-compatible commands over IEEE-488.2 interfaces, enabling seamless integration into complex test architectures.

## Technical Specifications

**Measurement Resolution & Speed**
• Resolution: 4.5 to 6.5 digits (selectable); up to 6.5 digits for DC voltage and resistance; 5.5 digits for AC current and AC voltage
• Read Rate: Up to 1000 readings/sec at 4.5-digit resolution; 5 readings/sec at 6.5-digit resolution (DC voltage)
• Data Storage: 1 to 1000 readings per block at high read rates

**DC Voltage (DCV)**
• Ranges: 100 mV to 300 V (decade steps)
• Resolution: 1 µV (5.5 digits), 100 nV (6.5 digits)
• Input Impedance: 10 GΩ (0.1–10 V ranges); 10 MΩ (100–300 V ranges)
• CMRR: > 146 dB at DC; > 80 dB + NMRR at 1–60 Hz
• NMRR: > 54 dB at 50/60 Hz; adds 20 dB with filter enabled
• Overvoltage Protection: 300 V RMS (all ranges)
• Accuracy (90 days, 23°C ± 5°C): 100 mV range ± (30 ppmR + 6 ppmFS); 10 V range ± (20 ppmR + 2 ppmFS); 300 V range ± (30 ppmR + 3 ppmFS)

**True RMS AC Voltage (ACV)**
• Ranges: 100 mV to 300 V (decade steps)
• Resolution: 1 µV (5.5 digits)
• Input Impedance: 1 MΩ / 100 pF
• CMRR: > 80 dB (DC to 60 Hz)
• Crest Factor: 5:1 at full range
• Overvoltage Protection: 300 V RMS (all ranges)
• Accuracy (90 days, 23°C ± 5°C): 10–40 Hz ± (0.4%R + 0.1%FS); 40 Hz–20 kHz ± (0.035%R + 0.01%FS); 300 kHz–1 MHz ± (2%R + 1%FS)

**Resistance (Ω)**
• Ranges: 100 Ω to 10 MΩ (decade steps)
• Resolution: 10 µΩ (6.5 digits)
• Maximum Lead Resistance: 100 Ω (any or all leads)
• Accuracy (90 days, 23°C ± 5°C): 100 Ω range ± (35 ppmR + 6 ppmFS); 10 kΩ range ± (30 ppmR + 3 ppmFS); 10 MΩ range ± (200 ppmR + 4 ppmFS)

**Integration & Control**
• VXI Conformance: VXI specification revision 1.4
• Programming: SCPI-compatible, IEEE-488.2
• Triggering: Front-panel or VXI backplane (synchronous and asynchronous)
• Calibration: Programmable external calibration (covers-on)

## Typical Applications

Automated component testing, precision voltage and resistance validation, high-speed data acquisition in production test environments, and integration into rack-mounted VXI test systems requiring traceable measurement standards.

## Compatibility & Integration

The 1362S operates as a C-size VXI module compatible with standard VXIbus architectures. SCPI command language ensures cross-platform software integration, while dual triggering modes support synchronized multi-module measurement sequences.

MPN

1362S

Frequency Range

10 Hz to 1 MHz

Voltage / Current Range

100 mV to 300 V (DCV and ACV); 100 Ω to 10 MΩ (resistance)

Brand Name

Fluke/Wavetek/Datron

You May Also Like