Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

HP / Agilent 66309D DC Source 0-15V 0-3A

SKU: 66309DCategories: DC Power Supplies
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The HP/Agilent 66309D is a DC power supply, providing a voltage output of 0-15V and a current output of 0-3A. It is suitable for a variety of electronic testing and development applications. This DC source offers reliable and stable power for your testing needs.

Request CalibrationRequest Repair
Product Starting From:

$790.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 66309D
Equipment info

The HP/Agilent 66309D is a dual-output DC power supply purpose-built for mobile communications and battery device testing. Each of its two independent outputs delivers 0–15 V at 0–3 A (45 W max per output, with 5 A peak capability for up to 7 ms). A built-in digital voltmeter provides direct measurement of external voltages from −4.5 Vdc to +25 Vdc, eliminating the need for external instrumentation.

## Technical Specifications

Output ratings span 0–15 V and 0–3 A per channel with 45 W maximum power per output. Peak current reaches 5 A for transient testing up to 7 ms. Programming accuracy is 0.05% + 10 mV for voltage and 0.05% + 1.33 mA for current (at 25°C ±5°C). DC readback accuracy for voltage measures 0.03% + 5 mV. Current readback ranges from 0.2% + 0.5 mA (positive range) to 0.1% + 2.5 µA in the ±20 mA range.

Voltage noise and ripple stay exceptionally low: 1.0 mV RMS and 6 mV peak-to-peak. Current ripple is 2 mA RMS. Transient response time is less than 35 µs with a 70 mV voltage drop. Output programming settles in under 200 µs rise/fall time, with full settling to 1 LSB within 6 ms (2 ms in fast mode).

The integrated DVM achieves 0.04% + 5 mV DC accuracy and 1% + 5 mV AC+DC RMS accuracy (60 Hz to 10 kHz). Common mode measurement range extends to −4.5 Vdc to +25 Vdc.

## Key Features

• Simultaneous independent control of main battery and charger circuitry testing
• Excellent voltage transient response prevents test interruptions from wiring voltage drops
• Open-sense wire connection detection for cellular telephone applications
• Adaptive fan speed for quiet operation at reduced power levels
• Built-in digital voltmeter eliminates external measurement equipment
• GPIB and remote interfaces for automated test systems

## Typical Applications

Mobile device battery characterization, wireless charging circuit validation, cellular telephone power integrity testing, and simultaneous battery/charger circuit evaluation.

## Compatibility & Integration

GPIB connectivity enables integration into automated test benches. Optional 521 adds solid-state output disconnect relays. Rackmount kits available: Option 1CM (standard) and Option AXS (side-by-side). Weight: 9.7 kg. Operating altitude to 2000 m, humidity to 95%.

MPN

66309D

Voltage / Current Range

0–15 V, 0–3 A per output; 5 A peak (up to 7 ms)

Brand Name

HP / Agilent