The Keithley 2260B-250-13 is a 1080W programmable DC power supply engineered for benchtop and system-level applications demanding precise voltage and current control. This single-output instrument delivers 0–250V and 0–13.5A with programmable rise/fall times, constant current priority mode, and simulated battery output characteristics via adjustable internal resistance. Fast 1ms transient response and comprehensive protection (OVP, OCP, OTP) make it suitable for component testing, production validation, and design verification across research, quality control, and manufacturing environments. The unit accommodates both analog control via 26-pin connector and digital interfaces for laboratory and automated test system integration.
Technical Specifications
• Output Voltage: 0 V to 250 V
• Output Current: 0 A to 13.5 A
• Power Rating: 1080W (1.08kW)
• Voltage Slew Rate: 0.1 V/s to 60 V/s
• Current Slew Rate: 0.01 A/s to 144 A/s
• Transient Recovery Time: 1ms
• Programmable Internal Resistance: Up to 18.51Ω
• Overvoltage Protection: 10% to 100% of rated output voltage (±2% accuracy)
• Overcurrent Protection: ~10% to 110% of rated output current (±2% accuracy)
• Over-Temperature Protection: Activated at elevated internal temperatures
• Isolation Voltage: 500 VDC (output to chassis)
• Output Connectors: IEC 320 standard input connector
– Key Features
• Programmable rise and fall times prevent inrush current damage to low-impedance loads
• Constant current priority setting reduces voltage and current overshoot for LED testing
• Internal test sequence mode stores and repeats voltage levels over specified time intervals
• Fast discharge capability rapidly dissipates output voltage
• Simulated battery output with variable internal resistance for multiple chemistry testing
– Typical Applications
• LED and semiconductor device characterization
• Battery simulation and load testing
• Component stress and reliability verification
• Production test environments requiring multi-level waveform generation
• Design validation requiring transient load response analysis
– Compatibility & Integration
• Analog control via 26-pin connector for external voltage/resistance control and status monitoring
• USB interface for digital control and data acquisition
• Rack-mountable form factor for integrated test system deployment
• Measurement Category II (IEC 60664) rated for local AC mains protection