The LeCroy WaveJet 322 is a 2-channel, 200 MHz digital oscilloscope engineered for field service, debugging, and benchtop measurement tasks. Its portable design, rapid sub-3-second startup, and intuitive Auto Setup reduce setup time and operator training overhead. The 7.5-inch color VGA display provides sufficient real estate for waveform visualization alongside measurements and menu controls. With 1 GS/s single-shot sampling (2 GS/s interleaved), 500 kpts memory per channel, and 100 GS/s equivalent-time acquisition, the instrument captures both transient and repetitive signals with 1.75 ns rise time resolution. Peak detect circuitry with 1 ns period ensures glitch capture at full bandwidth.
Technical Specifications
• Bandwidth: 200 MHz with selectable 20 MHz limiting filter
• Channels: 2 analog inputs
• Sample Rate: 1 GS/s (all channels), 2 GS/s (interleaved), 100 GS/s (equivalent-time)
• Memory: 500 kpts per channel
• Vertical Range: 2 mV/div to 10 V/div (1 MΩ input); 2 mV/div to 2 V/div (50 Ω input)
• Gain Accuracy: ±(1.5% + 0.5% FS)
• Rise Time: 1.75 ns
• Vertical Resolution: 8 bits
• Input Impedance: 1 MΩ ±1.5% (20 pF || 16 pF) or 50 Ω ±1.5%
• Input Coupling: GND, DC 1 MΩ, AC 1 MΩ, DC 50 Ω
• Maximum Input: 400 V CAT I (1 MΩ); 5 Vrms (50 Ω)
• Timebase: 500 ps/div to 50 s/div (1 ns/div minimum in standard modes)
• Timebase Accuracy: 10 ppm (typical)
• Triggering: Edge, glitch, period, pulse width, pulse count, TV with auto/normal/single/stop modes
• Capture Time: 500 µs at 1 GS/s; 250 µs at 2 GS/s
– Key Features
• 26 automatic measurements including rise/fall time, frequency, period, RMS, overshoot
• Math functions: sum, difference, product, FFT (to 8 kpts)
• Replay mode stores 1024 waveform acquisitions
• 6-digit frequency counter
• Probe sense ring for automatic attenuation detection
• PP010 probes (1 per channel) included
– Interfaces & Connectivity
• USB host and device ports
• GPIB interface (optional)
• 10/100Base-T Ethernet (optional)
• BNC input connectors with probe sense detection
– Power
• 100–240 Vrms AC, 50/60 Hz (±5%); 100–120 Vrms at 400 Hz (auto-adaptive)
– Physical
• Dimensions: 190 mm (H) × 285 mm (W) × 102 mm (D)