The NI PCI-6034E is a multifunction data acquisition device for PCI bus systems, delivering 16-channel analog input with 16-bit resolution at up to 200 kS/s, plus digital I/O and counter/timer functionality. Designed for laboratory automation, research, design verification, and manufacturing test applications, this E Series card combines configurable input ranges, programmable gain amplification, and dual 24-bit counters in a compact form factor. Integration with NI-DAQmx driver software and RTSI bus timing synchronization enables straightforward system implementation across diverse measurement scenarios.
Technical Specifications
Analog Input
• 16 single-ended or 8 differential channels (software-selectable per channel)
• 16-bit resolution; effective resolution to 18 bits with dithering
• Maximum sampling rate: 200 kS/s
• Input ranges: ±10 V, ±5 V, ±0.5 V, ±0.05 V
• Maximum working voltage: ±11 V with respect to ground (Installation Category I)
• Input impedance: 100 GΩ in parallel with 100 pF
• Programmable gain instrumentation amplifier (NI-PGIA) for guaranteed settling times at all gains
Digital I/O
• 8 lines, individually programmable for input or output
• 5 V TTL/CMOS compatible
• Source: 13 mA at (VCC – 0.4) V; Sink: 24 mA at 0.4 V
• VCC + 0.5 V protection; high impedance at startup and reset
Counter/Timers
• Two 24-bit counter/timers
• 20 MHz clock frequency
• Supports encoder inputs (non-quadrature, count and direction) and PWM outputs
Physical & Environmental
• PCI bus interface; 68-pin SCSI-II connector
• Dimensions: 6.9 in. × 4.2 in.; Weight: 5.3 oz
• Operating temperature: 0 to 55 °C
• Storage temperature: –20 to 70 °C
• Recommended warm-up: 15 minutes
• +5 V internal supply with 1 A fuse protection
– Key Features
• RTSI bus for real-time synchronization across multiple boards
• Compatible with SCC and SCXI signal conditioning modules for expanded sensor capabilities
• Dual-use counter channels: operate as encoder inputs or PWM outputs (not simultaneously)
• High input impedance minimizes loading effects
– Typical Applications
Laboratory automation, research instrumentation, design verification testing, manufacturing test systems, and sensor data logging requiring multi-channel analog acquisition with digital timing control.
– Compatibility & Integration
Operates with NI-DAQmx driver software. Integrates with RTSI bus timing architecture and SCC/SCXI signal conditioning ecosystems for enhanced measurement flexibility.