The Pacific Power 3060-MS is a solid-state AC power source delivering 62.5 kVA / 50 kW per chassis through double conversion architecture—AC to DC rectification followed by high-frequency PWM inverter synthesis. This design enables precise AC power simulation with <1% voltage THD at 50/60 Hz and <2% at 400 Hz, along with ±0.2% voltage accuracy and ±1% load regulation. The unit operates across 0–120/208 V three-phase (0–354 V L-N, 0–708 V L-L true RMS) with configurable WYE or Delta connections and supports external transformer options to 600 V L-L. Frequency coverage spans 47–500 Hz standard, with optional extended range to 1000 Hz. Current delivery reaches 175 A RMS continuous per phase, 350 A pulse per phase for non-linear loads. Transient response completes in 150 µs (50% step) or 300 µs (100% step), recovering to ±3% within one-tenth cycle at 400 Hz. Overload capacity supports 110% for 1 hour, 125% for 10 minutes, or 150% for 10 seconds. Multi-chassis parallel architecture scales to 625 kVA. Universal input transformer provisions accept 208, 240, 380, and 480 VAC three-phase ±10%, eliminating input power factor constraints.
Technical Specifications
Output Power & Voltage
• 62.5 kVA / 50 kW per chassis
• 0–120/208 V three-phase nominal; 0–354 V L-N / 0–708 V L-L true RMS
• WYE or Delta configuration
• Voltage resolution: 0.1 VAC (front panel), 0.001 VAC (remote interface)
• Voltage accuracy: ±0.2% of range + calibration reference
Current
• 175 A RMS continuous per phase
• 350 A pulse per phase for non-linear loads
• Current resolution: 0.01 A (front panel), 0.001 A (remote interface)
• Current accuracy: ±0.2% of range + calibration reference
Frequency
• Fixed: 50 Hz, 60 Hz, 400 Hz
• Variable: 47–500 Hz standard; optional 20–1000 Hz with SCU/UPC32 option
• Digital oscillator control
Regulation & Performance
• Load regulation: ±1% @ 50/60 Hz, ±2% @ 400 Hz with AGC
• Voltage THD: <1% @ 50/60 Hz, <2% @ 400 Hz
• Transient response: 150 µs (50% step), 300 µs (100% step); ±3% recovery within <1/10 cycle @ 400 Hz
• Overload: 110% (1 hour), 125% (10 minutes), 150% (10 seconds)
Load Support
• Any power factor; full rated kVA delivered at all power factors
– Key Features
• Double conversion topology for output fidelity
• Ultra-low voltage distortion: <1% THD at industrial frequencies
• High current pulse capability for inductive and non-linear test loads
• Scalable to 625 kVA via multi-chassis parallel operation
• External transformer options: up to 600 V L-L
• Universal input provisions: 208, 240, 380, 480 VAC three-phase ±10%
– Typical Applications
• Product power supply testing and validation
• Power conversion equipment evaluation
• Industrial and laboratory AC source replacement
• Facility power simulation and R&D testing
• Non-linear load characterization
– Compatibility & Integration
• Standalone or multi-cabinet parallel configuration
• User interface for manual operation
• Remote interface with 0.001 V / 0.001 A resolution for automated test integration
• SCU/UPC32 option extends frequency range to 20–1000 Hz