Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Rigol RP1100D – High Voltage Differential Probe, DC-100 MHz, 7000 Vpp

SKU: RP1100DCategories: Laboratory Measurement Instruments
Brand: Rigol
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Rigol RP1100D is a high-voltage differential probe with a bandwidth of DC to 100 MHz and a maximum differential voltage of 7000 Vpp. It is designed for measuring high-voltage signals in a variety of applications, including power electronics, motor drives, and high-voltage power supplies.

Request CalibrationRequest Repair
Product Starting From:

$0.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote RP1100D
Equipment info

The Rigol RP1100D is a high-voltage differential probe engineered for precision measurement of high-voltage signals up to 7000 Vpp across DC to 100 MHz bandwidth. Built for power electronics, motor drives, and high-voltage power supply development, this probe delivers ±2% measurement accuracy with selectable attenuation ratios (100:1, 200:1, 500:1, 1000:1) to match varying signal levels. Its double-insulated design achieves CAT III safety rating at 6500 Vrms maximum voltage to ground, ensuring reliable operation in demanding test environments.

Technical Specifications

Measurement Range & Accuracy
• Maximum differential input voltage: 7000 Vpp (7 kV)
• Input voltage range: 700 Vpp to 7000 Vpp depending on attenuation setting
• Maximum voltage to ground: 6500 Vrms
• Measurement accuracy: ±2%

Frequency & Timing Performance
• Bandwidth: DC–100 MHz (−3 dB) at ×200, ×500, ×1000 attenuation
• Bandwidth: DC–50 MHz (−3 dB) at ×100 attenuation
• Rise time: 3.5 ns (×200, ×500, ×1000); 7 ns (×100)

Signal Characteristics
• Attenuation ratios: 100:1, 200:1, 500:1, 1000:1
• Input impedance (differential): 16 MΩ / 1.2 pF or 100 MΩ / 1.2 pF
• Input impedance (single-ended/ground): 50 MΩ / 2.3 pF
• Output impedance: 50 ohms
• Output voltage: ≤ ±8.0 V
• CMRR: >80 dB @ 60 Hz, >60 dB @ 100 Hz, >50 dB @ 1 MHz

Power & Physical
• Power supply: 9V DC external
• Power consumption: <300 mA, 0.4 watt
• Dimensions: 240 mm × 85 mm × 36 mm
• Weight: 280 g
• Operating temperature: 0–50°C

Safety & Standards
• IEC 1010-1, CAT III, pollution degree 2
• Double insulation
• EN50081-1 and EN50082-1 compliance

– Key Features

• Four switchable attenuation settings optimize dynamic range for diverse voltage levels
• High common-mode rejection ratio isolates differential signals from ground noise
• 100 MHz bandwidth at highest attenuation preserves signal fidelity across wide frequency spectrum
• Dual-banana plug and BNC connectors enable flexible test setup configurations
• Compact form factor (280 g) supports portable field measurements

– Typical Applications

• Power electronics and inverter circuit analysis
• Motor drive signal characterization
• High-voltage power supply validation and troubleshooting
• Differential voltage measurement in floating circuits

– Compatibility & Integration

The RP1100D interfaces with all Rigol oscilloscopes and general-purpose oscilloscopes via standard BNC connectors. Included accessories—dual-BNC coax cable (100 cm), high-voltage silicon cable (60 cm), HV-IC-clips, alligator clips, and contact probe prods—support immediate deployment across laboratory and field environments.

MPN

RP1100D

Frequency Range

DC to 100 MHz (-3 dB) at x200, x1000 attenuation; DC to 50 MHz (-3 dB) at x100 attenuation, x500

Voltage / Current Range

7000 Vpp maximum differential input; 700 Vpp to 7000 Vpp input voltage range depending on attenuation; 6500 Vrms maximum to ground; ±8.0 V output; <300 mA power consumption

Brand Name

Rigol

You May Also Like

Can't Find The Right Product? Contact Us.