Call +1 ‪(484) 841-9341‬

NIST TRACEABILITY 

NIST-traceable calibration certificate available with data

Service Support

Service plans available for the products’ durability

FAST REPAIRS

1–3 day turnaround, 15-day return on repair policy

Software Updates 

Staying up-to-date with latest software updates.

Rod-L M100BVS5-2.8-40 AC Hipot Test Instrument

SKU: M100BVS5-2.8-40Categories: Hi-Pot Testers
Brand: Rod-L
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Rod-L M100BVS5-2.8-40 AC Hipot Test Instrument is designed for performing high voltage AC tests. It features a 5 kV AC output, a 2.8 mA trip current, and a 40 VA output. Ideal for ensuring electrical safety and insulation integrity in various electronic components and systems.

Request CalibrationRequest Repair
Product Starting From:

$410.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote M100BVS5-2.8-40
Equipment info

The Rod-L M100BVS5-2.8-40 is an AC hipot test instrument engineered for dielectric withstand testing of electrical equipment and components. It verifies insulation integrity by applying high AC voltage stress while monitoring leakage current and ground continuity. Designed for laboratory and field use, the M100BVS5-2.8-40 meets UL, IEC, CSA, VDE, and FCC standards and was the first hipot tester line to achieve UL Listing.

Technical Specifications

Electrical Performance
• Output voltage: 0 to 2.8 kV AC variable
• Leakage current range: up to 40 mA maximum
• Output power: 40 VA
• Test duration: 1 to 90 seconds adjustable
• Voltage ramp: Electronically controlled with smooth, consistent rise time
• High voltage shutdown: Within 2 ms of fault detection
• Power input: 115/230 VAC ±10%, 44–66 Hz (configured for 120 V AC)

Metering & Display
• Analog meters for voltage and current readout
• Digital panel meters available as option
• Failure indicator lamp
• Audible and visible alarm outputs for fault conditions (manual reset)

– Key Features

• Adjustable trip current with selectable front panel control options
• Dual fully automatic test cycles with electronic voltage ramp selection
• Ground continuity monitoring before and during test execution
• Automatic high voltage shutdown within 2 ms of breakdown detection
• Rear panel leakage current limit adjustment
• Front panel HV connector and DUT outlet for direct component connection

– Safety & Compliance

Operator protection design with rapid fault response. Meets UL, IEC, CSA, VDE, and FCC regulatory requirements for electrical equipment safety testing.

– Typical Applications

Dielectric withstand verification for electronic components, assemblies, and finished systems in manufacturing, quality assurance, and field maintenance environments.

MPN

M100BVS5-2.8-40

Voltage / Current Range

0–2.8 kV AC output, 120 VAC input, 40 mA maximum leakage current

Brand Name

Rod-L

You May Also Like