The Rod-L M100BVS5-2.8-40 is an AC hipot test instrument engineered for dielectric withstand testing of electrical equipment and components. It verifies insulation integrity by applying high AC voltage stress while monitoring leakage current and ground continuity. Designed for laboratory and field use, the M100BVS5-2.8-40 meets UL, IEC, CSA, VDE, and FCC standards and was the first hipot tester line to achieve UL Listing.
Technical Specifications
Electrical Performance
• Output voltage: 0 to 2.8 kV AC variable
• Leakage current range: up to 40 mA maximum
• Output power: 40 VA
• Test duration: 1 to 90 seconds adjustable
• Voltage ramp: Electronically controlled with smooth, consistent rise time
• High voltage shutdown: Within 2 ms of fault detection
• Power input: 115/230 VAC ±10%, 44–66 Hz (configured for 120 V AC)
Metering & Display
• Analog meters for voltage and current readout
• Digital panel meters available as option
• Failure indicator lamp
• Audible and visible alarm outputs for fault conditions (manual reset)
– Key Features
• Adjustable trip current with selectable front panel control options
• Dual fully automatic test cycles with electronic voltage ramp selection
• Ground continuity monitoring before and during test execution
• Automatic high voltage shutdown within 2 ms of breakdown detection
• Rear panel leakage current limit adjustment
• Front panel HV connector and DUT outlet for direct component connection
– Safety & Compliance
Operator protection design with rapid fault response. Meets UL, IEC, CSA, VDE, and FCC regulatory requirements for electrical equipment safety testing.
– Typical Applications
Dielectric withstand verification for electronic components, assemblies, and finished systems in manufacturing, quality assurance, and field maintenance environments.