The VITREK 944i is a high-performance dielectric analyzer used for characterizing the dielectric properties of materials. It measures capacitance, dissipation factor (tan δ), and impedance over a wide frequency range. Ideal for quality control, R&D, and materials science applications.

Request CalibrationRequest Repair
Product Starting From:

$2,630.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 944i
Equipment info

The VITREK 944i is a multi-functional dielectric analyzer that combines AC Hipot, DC Hipot, insulation resistance, and earth continuity testing in a single compact instrument. Purpose-built for manufacturing test, product safety engineering, component evaluation, and R&D environments, the 944i delivers four critical electrical safety measurements from one connection point. Its automated, menu-driven interface and compliance with UL, CSA, IEC, and TUV standards make it suitable for high-performance applications requiring simultaneous verification of dielectric strength and insulation integrity.

Technical Specifications

Dielectric Strength Testing
• AC Hipot: up to 5 kV AC standard, 35 mA sourcing capability, 50 Hz to 100 Hz output frequency with 1 Hz resolution, digitally synthesized low-distortion sine wave
• DC Hipot: up to 5 kV DC standard (optional upgrades to 7 kV DC and 10 kV DC), user-selectable slew rate from 10 V/s to 1999 V/s
• Current Limits: 35 mA (20 mA below 500 V); leakage current measurement to 1 nA; real, apparent, and total current sensing modes; arc detection selectable 3 to 30 microseconds with 10 ms shutdown
• Voltage Accuracy: ±70.7 VAC / ±53.0 VDC at 5 kV; tolerance equals square root of output voltage (e.g., ±10.0 VAC at 100 VAC)

Insulation Resistance (IR) Testing
• Measurement range: 1 MΩ to 10 TΩ with expanded ranges to 500 GΩ or 65 GΩ
• Test current: down to 10 nA at 5 kV DC

Earth Continuity Testing
• Full scale: 8.5 Ω maximum (including lead resistance), 1 mΩ resolution
• Test current: 50 mA
• Automatic lead compensation; accuracy ±0.3% of reading + 30 mΩ

Display & Control
• 40-character alphanumeric LCD; 1 V AC/DC voltage resolution
• Stores up to 200 test steps; automatic test abort on limit exceedance
• Illuminated HV indicator with independent output voltage monitor

Safety & Diagnostics
• Built-in automatic calibration and diagnostics
• Internal watchdog safety monitor

– Key Features

• Four electrical safety test parameters from single connection
• Digitally synthesized AC output waveform with selectable frequency and current limits
• Leakage current sensitivity to 1 nA; insulation resistance to 10 TΩ
• Configurable arc detection with sub-microsecond response
• Automatic two-wire lead resistance compensation for continuity testing

– Typical Applications

• Manufacturing production test of electrical assemblies and components
• Product safety certification and compliance verification
• Insulation system evaluation and qualification
• Research and development of high-voltage and protective systems

– Compatibility & Integration

The 944i meets UL, CSA, IEC, and TUV safety standards, supporting widespread integration into test facilities requiring multi-parameter dielectric evaluation.

MPN

944i

Frequency Range

50 Hz to 100 Hz

Voltage / Current Range

35 mA (20 mA below 500V); DC Hipot: up to 5 kV DC (optional to 7 kV/10 kV); Continuity: 50 mA; IR: test current down to 10 nA, AC Hipot: up to 5 kV AC