The Wandel & Goltermann OLP-150 is an optical power meter engineered for fiber optic installation, maintenance, and repair. It delivers precise power and loss measurements across wavelengths from 800 nm to 1700 nm, with display ranges varying by model variant. The instrument supports both modulated and unmodulated light signals, automatic wavelength detection, and dual-wavelength attenuation measurement via TWINtest mode. Field deployment is enabled through battery operation (dry or rechargeable) or AC line adapter power, while rugged construction and interchangeable fiber connectors suit both field and laboratory environments.
Technical Specifications
Wavelength Range: 800 nm to 1700 nm
Photodiodes: InGaAs photodiodes used in select models (e.g., OLP-16B) for optimized performance at 1310 nm and 1550 nm
Fiber Compatibility:
• OLP-5, OLP-6: Single-mode and multimode fibers up to 100 µm core diameter
• OLP-8: Single-mode fibers with 9 µm core diameter
• Interchangeable adapter system supports common fiber optic connectors
Display Ranges by Model:
• OLP-15A/B: −45 dBm to +20 dBm
• OLP-16A/B: −55 dBm to +15 dBm
• OLP-18A/B: −35 dBm to +26 dBm
• Resolution below −60 dBm: 0.1 dB
Operating Temperature: −10°C to +55°C
Storage/Transport Temperature: −40°C to +70°C
Measurement Uncertainty: Specified across operating temperature range (−5°C to +45°C after zeroing, ΔT = ±1 K); reference conditions: −20 dBm CW, 1310 nm ±1 nm, 23°C ±3°C, 45–75% RH
Power Supply: Dry or rechargeable batteries (NiCd, Mignon AA); AC line adapter (e.g., NT-20) with optional charging capability
– Key Features
• Power and loss measurements across broad wavelength range
• Automatic wavelength detection (AUTO-λ mode) for single-wavelength attenuation
• TWINtest mode for simultaneous dual-wavelength attenuation when paired with compatible sources (OLS-15, OLS-5, OLS-6)
• Modulation detection for audible fiber identification
• Measurement of both modulated and unmodulated optical signals
– Typical Applications
• CATV systems
• Optical amplifier characterization (EDFA)
• High-power laser measurement
• Link attenuation and insertion loss verification
– Compatibility & Integration
Designed for use with Wandel & Goltermann light sources and compatible test equipment. Interchangeable adapter architecture accommodates standard fiber optic connector systems.