The Astronics/C&H Technologies MA209 is a 100 MHz pulse generator engineered for precision signal generation across research, development, and production test environments. Implemented as an M-Module conforming to ANSI/VITA 12-1996, it integrates with VXI, VME, PCI, and cPCI carrier boards. The MA209 delivers fully programmable frequency, pulse width, delay, and amplitude control with exceptional timing accuracy and flexible operational modes including single pulse, continuous, burst, and double-pulse generation.
## Technical Specifications
**Pulse Frequency (Internal Clock)**
• Range: 0.093 Hz to 100 MHz
• Resolution: 0.093 Hz
• Accuracy: ±0.01%
• Temperature Coefficient: +17 ppm/°C typ.
**Pulse Width**
• Range: 5 ns to (period – 3 ns)
• Accuracy: ±(3% + 250 ps)
• Resolution: 10 ps to 20 ps depending on width; approximately 10 ps per 12 ns increment above 24 ns
**Pulse Delay**
• Range: 5 ns to 5 s
• Accuracy: ±(3% + 250 ps)
**Double Pulse Spacing**
• Range: (width + 3 ns) to (period – width – 3 ns)
• Accuracy: ±(3% + 250 ps)
**Pulse Output**
• Amplitude Range: -1.5 V to +6.5 V (open circuit)
• Output Impedance: 50 Ω
• Amplitude Resolution: 2 mV (12-bit)
• Amplitude Accuracy: ±(2.0% + 100 mV)
• Output Current: ±60 mA typical; ±120 mA max (short circuit, dynamic)
• Programmable Rise/Fall Time: 0.625 to 2.5 V/ns
**Sync Output Latency**
• Time from External Trigger: 1 billion pulses), and double-pulse modes
• Follow-trigger and inverted pulse/sync operation
• Dual external trigger inputs (front panel and M-Module)
• Gate, trigger, and reference clock inputs selectable from front panel or M-Module backplane
• Pulse and Sync outputs routable to front panel or M-Module backplane connectors
• ±0.01% frequency accuracy with disciplined external reference
## Typical Applications
Pulse characterization, radar and sonar testing, timing synchronization, and circuit validation in laboratory and production environments.
## Compatibility & Integration
M-Module form factor with front-panel connectors (Pulse Output, Sync Output, Input A, Input B) and M-Module backplane Trig A & B signals. Operates from +5V (1500 mA max) and +12V supplies.