The Exfo WA-7600 Optical Channel Analyzer is a high-performance benchtop instrument for absolute wavelength and power measurement in DWDM and ROADM systems. Using scanning Michelson interferometer-based Wavemeter technology, it simultaneously measures the wavelengths and individual powers of up to 256 optical signals with absolute uncertainty of ±0.2 ppm or ±0.3 pm. The analyzer automatically calculates OSNR and channel spacing while graphing wavelength and power trends over time.
Technical Specifications
Wavelength Measurement
• Range: 1270 nm to 1680 nm (178 THz to 236 THz)
• Absolute uncertainty: ±0.2 ppm or ±0.3 pm
• Display resolution: 0.0001 nm
• Spectral resolution: 30 pm
• Measurement accuracy: 3σ confidence level (≥99.6%)
• NIST traceability confirmed
Power Measurement
• Uncertainty: ±0.5 dB at ±30 nm from 1310 nm and 1550 nm
• Display resolution: 0.01 dB
• Input sensitivity: -40 dBm (0.1 µW) for 1270–1600 nm; -30 dBm (1.0 µW) for 1600–1680 nm
• Maximum input level (all lines): 10 dBm (10 mW)
• Maximum safe level (all lines): 18 dBm (63 mW)
Optical Dynamic Range
• >35 dB for channel spacing ≥100 GHz
• >27 dB for channel spacing ≥50 GHz
Measurement Performance
• Maximum laser lines: 256
• Cycle time: 1.25 s (0.8 measurements/s)
• Warmup time: 7 minutes nominal
– Key Features
• Discriminates 100 GHz, 50 GHz, and future 25 GHz channel spacing
• Simultaneous individual peak and total optical power measurement
• Automatic OSNR and channel spacing calculation
• Real-time wavelength and power trending
• Measurement accuracy verified at 3σ (≥99.6% confidence)
– Typical Applications
• DWDM system characterization and performance verification
• ROADM optical signal analysis
• Multi-channel wavelength and power profiling
• Channel spacing and OSNR monitoring
– Compatibility & Integration
Optical Input: Single mode 9/125 µm fiber with FC/UPC, FC/APC, SC/UPC, SC/APC, or ST/UPC connectors
Interfaces: GPIB (IEEE-488.2), RS-232, VGA monitor port, parallel printer port, floppy disk drive
Physical: 133 mm (H) × 432 mm (W) × 419 mm (D); rugged benchtop or rack-mounted configuration
Environment: Operating 10–30°C; storage -10–70°C; ≤90% RH at +40°C (non-condensing)