The Fluke 8050A-01 is a 4.5-digit digital multimeter designed for accurate and reliable electrical measurements. It features a clear display, multiple measurement functions (voltage, current, resistance), and durable construction. Ideal for electronics troubleshooting, maintenance, and calibration applications.

Request CalibrationRequest Repair
Product Starting From:

$79.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote 8050A-01
Equipment info

The Fluke 8050A-01 is a 4.5-digit digital multimeter engineered for precision electrical measurements across DC voltage, AC voltage, current, resistance, and specialized functions including diode testing and dBm measurement. This portable instrument delivers 0.03% basic DC accuracy with true RMS capability from 20 Hz to 50 kHz, making it suitable for electronics troubleshooting, field service, and calibration work. The model 8050A-01 integrates rechargeable NiCad batteries providing 10 hours typical operating time, alongside AC line power operation (90–264 V, 47–440 Hz), enabling both cordless field deployment and stationary bench use. An automatic polarity indicator and overrange display enhance readability on the 4.5-digit LCD. Built-in transient protection rated to 6 kV, 10 microseconds, combined with overload protection and a 500 V dc maximum common mode voltage rating, safeguards the instrument against typical field hazards. Weighing 1.08 kg and measuring 22 cm × 6 cm × 25 cm, the meter fits standard equipment racks and portable kits.

## Technical Specifications

**Measurement Functions & Ranges**
• DC Voltage: 10 µV to 1000 V
• AC Voltage: 10 mV to 750 V true RMS (linear and dB)
• DC Current: 10 nA to 2 A
• AC Current: 10 µA to 2 A true RMS
• Resistance: 10 mΩ to 20 MΩ
• Conductance: Up to 100,000 MΩ (200 nS)
• dBm Measurement: Selectable reference impedances

**Performance**
• Accuracy (DC): 0.03% basic
• True RMS bandwidth: 20 Hz to 50 kHz
• Operating temperature: 0°C to 50°C
• Storage temperature: −40°C to +70°C (without batteries); −40°C to +50°C (with batteries)
• Humidity: Up to 90% (0°C–35°C); up to 70% (35°C–50°C)
• Temperature coefficient: <0.1× specification per °C (0°C–18°C, 28°C–50°C)
• Calibration interval: 1 year

**Power & Protection**
• Rechargeable NiCad batteries: 10 hours typical operation; 14 hours recharge time (OFF position)
• AC line power: 90–264 V, 47–440 Hz (field changeable)
• Transient protection: 6 kV, 10 microseconds
• Maximum common mode voltage: 500 V dc or peak ac

## Key Features

• Diode and transistor testing via forward voltage measurement
• Relative reference function across all measurement modes
• Automatic polarity indication and overrange display
• True RMS AC measurement independent of waveform shape

## Typical Applications

Electronics diagnostics, field service maintenance, power supply validation, component characterization, and frequency-dependent AC measurements in industrial and laboratory environments.

## Compatibility & Integration

Model 8050A-01 includes rechargeable battery capability. Standard model 8050A operates on AC line power only. Dual input connectors (HI and COMMON) support standard test lead configurations.

MPN

8050A-01

Frequency Range

20 Hz to 50 kHz (true RMS AC bandwidth)

Voltage / Current Range

DC voltage: 10 µV to 1000 V; AC voltage: 10 mV to 750 V; DC current: 10 nA to 2 A; AC current: 10 µA to 2 A