The HP/Agilent 6282A is a regulated DC power supply delivering 0–10 VDC at 0–10 ADC with 100 W maximum output. Designed for laboratory, test, and system integration environments, it provides precise voltage and current control via front-panel potentiometers and dual digital displays. The unit combines tight load and line regulation, sub-millisecond response, and comprehensive protection into a compact 19-inch rack-mountable chassis.
Technical Specifications
Output
• Voltage: 0–10 VDC
• Current: 0–10 ADC
• Power: 100 W maximum
Regulation & Stability
• Load regulation (voltage): <0.01% + 2 mV (0–100% load current change)
• Load regulation (current): <0.02% + 2 mA (0–100% load voltage change)
• Line regulation (voltage): <0.01% + 1 mV (±10% line voltage change)
• Line regulation (current): <0.01% + 1 mA (±10% line voltage change)
• Ripple and noise (voltage): <1 mV rms
• Ripple and noise (current): <5 mA rms
• Response time (50% load step): <100 µs to 1% of final value
• Temperature coefficient (voltage): <0.015% + 0.5 mV/°C
• Temperature coefficient (current): <0.02% + 0.5 mA/°C
Metering
• Voltage display: Digital, 0–10 V range, ±(0.1% reading + 1 digit) accuracy
• Current display: Digital, 0–10 A range, ±(0.1% reading + 1 digit) accuracy
– Key Features
• Continuously variable voltage and current adjustment via potentiometers
• Built-in overvoltage protection (OVP) and overcurrent protection (OCP)
• Over-temperature protection circuitry
• Heavy-duty binding posts for banana plugs or bare wire
• Front-panel output terminals and metering
• Selectable input: 105–127 VAC or 210–250 VAC, 47–63 Hz
• Input current: ~4 A at 120 VAC
– Physical & Environmental
• Dimensions: 5.25 in H × 19 in W × 14.5 in D (133 × 483 × 368 mm)
• Weight: ~35 lbs (15.9 kg)
• Operating temperature: 0–50°C
• Storage temperature: -40–85°C
• Humidity: 20–80% RH, non-condensing
– Typical Applications
Powering sensitive electronic components, system integration testing, laboratory measurements, and bench-level characterization requiring stable, adjustable DC output with real-time voltage and current monitoring.