The Keysight/Agilent/HP 8082A is a dual-channel pulse generator delivering precise, variable pulse signals from 1 kHz to 250 MHz across six ranges. It generates pulses with widths as fast as 2.4 ns to 0.5 ms, adjustable delays from 2 ns to 0.5 ms, and transition times under 1 ns. The instrument produces outputs up to ±5 V with selectable ECL levels (-0.9 V to -1.7 V typical) and includes a 50 Ω source impedance matched output stage rated for short-circuit and open-circuit protection.
With period, width, and delay jitter specifications of <0.1% of setting +50 ps, the 8082A excels in digital circuit characterization, logic testing, and high-speed signal development. It supports normal self-contained operation, external trigger mode, gate mode, external width mode, and single/double pulse generation up to 125 MHz. Overshoot and ringing remain below 10% of pulse amplitude, with 5% preshoot. The dual three-step attenuators reduce outputs to 1 V, while a common vernier control extends minimum output to 0.4 V. An adjustable offset reaches +2 V into 50 Ω loads. Fixed delay between trigger and pulse output is approximately 17 ns.
Technical Specifications
• Repetition Rate: 1 kHz to 250 MHz (six selectable ranges)
• Pulse Width: 2.4 ns to 0.5 ms (six ranges)
• Pulse Delay: 2 ns to 0.5 ms (six ranges)
• Fixed Delay (trigger to output): Approximately 17 ns
• Period Jitter: <0.1% of setting +50 ps
• Width Jitter: <0.1% of setting +50 ps
• Delay Jitter: <0.1% of setting +50 ps
• Transition Times (leading/trailing edge): <1 ns
• Rise Time: 1 ns minimum
• Overshoot and Ringing: 5 ns transitions): <5% both slopes
• Amplitude: Up to ±5 V into 50 Ω load
• Offset: Up to +2 V into 50 Ω
• DC Source Impedance: 50 Ω ±5%–10%
• ECL Voltage Swing: -0.9 V to -1.7 V typical (open circuit)
• Output Protection: Withstands open circuit, short circuit, ±6 V, ±200 mA external application
– Key Features
• Dual independent channels with separate three-step attenuators
• Switch-selectable ECL output levels
• Vernier amplitude control to 0.4 V minimum
• Multiple operating modes: normal, external trigger, gate, external width
• Single and double pulse generation capability (double pulse: up to 125 MHz)
• Reflection coefficient: 5% ECL setting, 15% in 5 V range
– Typical Applications
• TTL and ECL logic circuit testing and validation
• Digital system characterization and development
• High-speed signal generation for research environments
• Pulse train development and timing analysis
– Compatibility & Integration
The 8082A features trigger output for synchronization with external instruments and accepts external trigger, gate, and width control signals for flexible test system integration.