The Kikusui PMC35-1A is a series-regulated DC power supply delivering 0 to 35 VDC at 0 to 1 A, engineered for laboratory and industrial electronics testing, product development, and quality control. This compact linear supply operates with constant voltage (CV) and constant current (CC) regulation modes, providing low-noise output suitable for precision measurement and device characterization. The front panel features a push-in ON/OFF switch with lock function, dual 10-turn potentiometers for voltage and current adjustment, and an LED digital meter displaying both voltage and current simultaneously. Load regulation performance reaches 0.5% + 1 digit for voltage and 1% + 1 digit for current, while line regulation holds at 0.1% + 1 digit and 0.5% + 1 digit respectively. Output ripple and noise are specified at 3 mVrms (voltage) and 10 mArms (current). The unit includes overvoltage protection (OVP) and overheating protection (OHP) as standard safety features. Dual input voltage configurations accommodate 105–130 VAC, 180–220 VAC, 195–239 VAC, or 210–250 VAC at 50/60 Hz. The rear panel provides a J2 connector for analog remote control and monitoring, fused AC input, and protective earth terminal. A 30-minute warm-up period is recommended to achieve specified performance stability.
## Technical Specifications
• Output voltage: 0 to 35 VDC
• Output current: 0 to 1 A
• Input voltage: 105–130 VAC, 180–220 VAC, 195–239 VAC, or 210–250 VAC
• Input frequency: 50/60 Hz
• Load regulation: 0.5% + 1 digit (voltage), 1% + 1 digit (current)
• Line regulation: 0.1% + 1 digit (voltage), 0.5% + 1 digit (current)
• Output ripple and noise: 3 mVrms (voltage), 10 mArms (current)
• Warm-up time: 30 minutes
• Dimensions (Type I case): 175 mm W × 83 mm H × 250 mm D
• Weight: 4.0 kg
## Key Features
• Constant voltage and constant current regulation modes
• LED digital meter with simultaneous voltage and current display
• 10-turn potentiometers for precise voltage and current control
• Push-in ON/OFF switch with mechanical lock
• Overvoltage and overheating protection
• Low output noise design
## Compatibility & Integration
• Analog remote control via rear-panel J2 connector
• GPIB integration when used with power supply controllers (e.g., PIA4800 Series)
• GPIB-controlled functions: output voltage/current setting and readback, power switch monitoring, CV/CC mode status, output state monitoring, and alarm monitoring