The Power Ten 163C is a 10 kW DC power supply engineered for semiconductor burn-in, process control, production testing, and laboratory research. It delivers stable, efficient power across more than 54 standard output voltage and current configurations using PWM switching technology with zero crossing topology. This unit accepts single-phase or three-phase AC input (190–253 VAC standard, with optional 360–440 VAC and 432–528 VAC ratings) and maintains exceptional regulation and thermal stability across demanding industrial and laboratory environments.
Technical Specifications
Power Output: 10 kW maximum
Input Power:
• Standard: 190–253 VAC, 47–63 Hz
• Optional: 360–440 VAC, 47–63 Hz
• Optional: 432–528 VAC, 47–63 Hz
• Input current @ 208 VAC: 15 A (3.3 kW), 28 A (6.6 kW), 41 A (10 kW)
• Delta or Wye source compatible
Output Voltage & Current: Over 54 standard combinations available. Representative configurations: 15 V/660 A, 0–30 V/330 A, 10 V/1000 A, 20 V/166 A
Regulation:
• Voltage: ±0.1% of maximum rated output
• Current: ±0.5% of maximum rated output
Stability: ±0.05% of set point per 8 hours (30 minutes warm-up, fixed line/load/temperature)
Transient Response: 30% current step recovers to ±2% of set voltage within 10 ms
Output Ripple: 50 mV P–P typical (model dependent, 30 mV max on select versions)
Temperature Coefficient:
• Voltage: 0.02%/°C of rated output voltage
• Current: 0.03%/°C of rated output current
– Key Features
• High-efficiency PWM switching with zero crossing topology
• Power factor: 0.99 min. at full load
• Total harmonic distortion: 2.5% max
• Bus bar output interface with 0.390″ holes (≤50 VDC)
• DB 25-pin control/monitor connector
• Remote on/off via contact closure (6–120 VDC, TTL/CMOS)
• Output voltage and current monitoring
• OVP limit adjustment
• Summary fault status indication
– Typical Applications
• Semiconductor burn-in testing
• Production line test systems
• Process control equipment
• Electronic research laboratories
– Compatibility & Integration
AC input terminal: 3 wire + GND (#10-32 threaded studs) for 208 VAC configurations. Remote programming and monitoring via rear-panel DB 25-pin female connector.