Tektronix TIVP05 IsoVu 500 MHz Isolated Probe with 2 m cable

SKU: TIVP05Categories: Laboratory Measurement Instruments
Brand: Tektronix
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Tektronix TIVP05 IsoVu™ is a galvanically isolated probe capable of measuring differential signals up to ±2500 V with a bandwidth of 500 MHz. It features a 2-meter cable and provides high common-mode rejection for accurate measurements in high-voltage environments.

Request CalibrationRequest Repair
Product Starting From:

$20,970.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote TIVP05
Equipment info

The Tektronix TIVP05 IsoVu is a 500 MHz galvanically isolated differential probe engineered for high-voltage, high-bandwidth measurements in power electronics. Using optical isolation via power-over-fiber and optical analog signal transmission, it achieves exceptional common-mode rejection and enables accurate differential measurements across ±2500 V with common-mode voltages reaching ±60 kV peak. This second-generation design combines a compact form factor—one-fifth the size of first-generation probes—with superior accuracy and sensitivity for demanding power device analysis.

Technical Specifications

• Bandwidth: 500 MHz with 850 ps rise time
• Differential Input Voltage Range: ±2500 V
• Common Mode Voltage Range: ±60 kV peak
• Common Mode Rejection Ratio: 160 dB DC; 100 dB at 100 MHz and 200 MHz; up to 80 dB at 1 GHz
• Input Capacitance: < 2 pF
• Differential DC Gain Accuracy: < 2% (typically < 1.5% after self-calibration)
• Offset Range: Up to ±2500 V
• Slew Rate Capability: Up to ±60 kV at 100 V/ns or faster
• Cable Length: 2 meters with MMCX connectors

– Key Features

• Galvanic isolation through fiber optic connection eliminates ground loops and reduces common-mode interference
• Solid metal probe body shields center conductor, minimizing ground loop area
• Compact probe head design enables access in space-constrained test environments
• Shielded coaxial cable with integrated fiber optic connection
• Offset capability supports floating measurements and multi-level switching topologies

– Typical Applications

• SiC and GaN FET bridge designs (half and full bridge)
• IGBT circuit evaluation
• Power converter and switching power supply analysis
• Motor drive and inverter design
• Electronic ballast circuits
• Wide bandgap device characterization
• EMI and ESD troubleshooting
• Current shunt and BCI measurements

– Compatibility & Integration

Exclusively compatible with Tektronix 4, 5, and 6 Series MSO oscilloscopes. Standard accessory kit includes IsoVu carrying case, wrench/driver, probe tip adapter, bipod, TIVPMX10X MMCX probe tip, quick reference guide, and traceable calibration certificate. Additional probe tip adapters available for specialized connection requirements.

MPN

TIVP05

Frequency Range

DC to 500 MHz

Voltage / Current Range

±2500 V differential, ±60 kV common mode peak