The Xantrex XFR100-28 is a 2800 W programmable DC power supply engineered for precision output regulation across research, development, and production testing environments. Built on zero-voltage soft-switching topology, the unit minimizes switching transients to achieve 81% efficiency, reduced thermal output, and extended operational life. The supply accepts single-phase 190–264 VAC input at 47–63 Hz and delivers 0–100 Vdc at 0–28 A with exceptional load and line regulation.
Technical Specifications
• Output Power: 2800 W
• Output Voltage Range: 0–100 Vdc
• Output Current Range: 0–28 A
• Input: 190–264 VAC, 1-phase, 47–63 Hz; 22.6 A @ 208 VAC, 20.5 A @ 230 VAC (typical)
• Output Noise: 175 mVp-p (0–20 MHz)
• Output Ripple: 25 mV voltage (rms), 80 mA current (rms)
• Load Regulation: 27 mV voltage, 10.6 mA current
• Line Regulation: 4 mV voltage (typical), 4.8 mA current (typical)
• Voltage Drift: 3.8 mV over 8 hours; 1.5 mV/°C temperature coefficient
• Efficiency: 81% typical @ 200 VAC and rated output
• Switching Frequency: 31 kHz nominal (62 kHz output ripple)
• Transient Response: <3 ms for voltage recovery within 0.5% after step change
• Stabilization Time: 7 s maximum from power-on
• Remote Sense Compensation: 5 V line loss
– Key Features
• Over-voltage protection (OVP) adjustable 5–110% of maximum voltage
• Over-temperature and thermal shutdown protection standard
• Dual 10-turn front-panel controls with simultaneous voltage/current display
• Constant voltage and constant current operation modes
• Remote analog programming: voltage and current (0–5 kΩ, 0–10 kΩ isolated sources)
• Remote analog monitoring: 0–5 V and 0–10 V output ranges (0–100% scaling)
• Remote on/off control with selectable TTL or 2.5–15 V logic
• Standby mode and remote sensing capability
– Typical Applications
Laboratory power conditioning, industrial test benches, product qualification cycles, and production burn-in environments requiring stable, regulated DC output with low noise and fast transient response.
– Compatibility & Integration
Density-optimized design permits rack-mount stacking without inter-unit ventilation gaps. Single-phase mains compatibility suits standard laboratory and industrial installations.