The HP/Agilent E3632A is a 120 W single-output DC power supply engineered for benchtop and system integration. It delivers dual output ranges—0 to 15 V at 7 A, or 0 to 30 V at 4 A—making it suitable for diverse power delivery requirements. The unit combines low output ripple and noise with excellent line and load regulation, backed by built-in digital measurement and programmable protection circuitry.
Technical Specifications
Output Ratings
• Range 1: 0–15 V @ 0–7 A
• Range 2: 0–30 V @ 0–4 A
• Total Output Power: 120 W
Programming Accuracy (25°C ±5°C)
• Voltage: ±0.05% of output + 10 mV
• Current: ±0.2% of output + 10 mA
Programming Resolution
• Voltage: 1 mV
• Current: 0.5 mA
Readback Accuracy (25°C ±5°C)
• Voltage: ±0.05% of output + 5 mV
• Current: ±0.15% of output + 5 mA
Readback Resolution
• Voltage: 0.5 mV
Ripple and Noise (20 Hz to 20 MHz)
• Normal-mode Voltage: <350 µVrms / 2 mVpp
• Normal-mode Current: <2 mA rms
• Common-mode Current: <1.5 µA rms
Regulation
• Line Regulation: 0.01%
• Load Regulation: 0.01%
Protection
• Overvoltage Protection: Programmable 1–32 V trip, ±0.5% + 0.5 V accuracy
• Overcurrent Protection: Programmable trip current level
Transient Response: <50 µs recovery to within 15 mV following full-load to half-load transitions.
Command Processing: <100 ms output voltage response via GPIB or RS-232.
– Key Features
• Dual output ranges maximize flexibility across low and high voltage applications
• Remote sensing supported for load-end voltage regulation
• Constant Voltage and Constant Current output modes
• SCPI-compliant programming via GPIB (IEEE-488) and RS-232
• Three programmable memory slots for setup storage and recall
• Front-panel control with dedicated knob and selection keys
• Compact benchtop form factor: 213 mm (W) × 133 mm (H) × 348 mm (D); 10.75 kg
– Typical Applications
Laboratory test and characterization, automated system test, component validation, and general-purpose DC power delivery in environments requiring both high voltage and high current capability.
– Compatibility & Integration
Interfaces: GPIB, RS-232. Input voltage selection: 100 VAC ±10%, 115 VAC ±10%, or 230 VAC ±10% (47–63 Hz). Operating temperature: 0°C to 40°C.