The Sorensen XTS15-4 is a 60 W programmable linear DC power supply for benchtop and system integration. It delivers 0–15 V and 0–4 A with line regulation of 0.02% (voltage) and 0.03% (current), making it suitable for electronics testing, circuit prototyping, and educational lab work. The unit operates in both constant voltage and constant current modes with automatic crossover, eliminating the need for manual switching during load transitions.
Technical Specifications
Output Performance
• Output voltage: 0–15 V DC
• Output current: 0–4 A DC
• Output power: 60 W
• Load regulation: 0.02% + 4 mV (voltage); 0.02% of Imax (current)
• Output noise and ripple (20 Hz–20 MHz): Typically 6 mV (voltage); 0.6 mA (current)
• Program accuracy: 10 mV + 0.1% of set voltage; 110 mA + 0.15% of set current
• Readback accuracy: 10 mV + 0.15% of Vmax; 110 mA + 0.15% of Imax
• Transient response: <100 µs for ±50% load change (typical)
Stability
• Line regulation: 0.02% of Vmax (voltage); 0.03% of Imax (current)
• Drift (8 hours): 0.02% of Vmax (voltage); 0.03% of Imax (current)
• Temperature coefficient: Typically 4.5 mV/°C (voltage); 0.4 mA/°C (current)
– Key Features
• Simultaneous digital LED displays for voltage and current readout
• Twin LED analog bar graphs showing relative voltage and current levels
• Overvoltage protection (OVP) and overcurrent protection (OCP) with status indication
• No overshoot at turn-on, turn-off, or power failure
• Impedance-switched remote sensing capability
• Series, parallel, and split-supply operation modes supported
• Optional IEEE-488 (GPIB) interface for automated testing
• Compact quarter-rack package; rack adapters available
– Typical Applications
• Electronics testing and verification
• Circuit prototyping and breadboarding
• Educational laboratory use
• System-level power distribution in multi-unit configurations
– Compatibility & Integration
The XTS15-4 can be configured as single units or combined into dual (XTD), triple (XTT), and quad (XTQ) packages. Optional GPIB control enables remote programming and monitoring. Input power: 115 VAC, 60 Hz single phase (optional inputs available).