Teledyne LeCroy HVD3102A-NOACC High Voltage Differential Probe, 1 kV, 25 MHz

SKU: HVD3102A-NOACCCategories: Laboratory Measurement Instruments
30 Days Warranty30 Days Free Returns >

+1 ‪(484) 841-9341‬

  • Repair Support & OEM Spare Parts
  • University and Research Discounts
  • Rental Units Available

The Teledyne LeCroy HVD3102A-NOACC is a high-voltage differential probe designed for measuring high-voltage signals with accuracy and safety. It features a voltage range of up to 1 kV and a bandwidth of 25 MHz, making it suitable for a variety of applications. This probe allows for safe and accurate measurements of high-voltage signals in circuit design, power electronics, and other high-voltage environments.

Request CalibrationRequest Repair
Product Starting From:

$2,729.00

In stock

In stock

Credit cards Accepted, Visa, ,mastercard, discover, PAypal, amex
Precision RF and optical test equipment sales, calibration, and repair by Aumictech. Email: sales@aumictech.com

Fastest Shipping. Estimated 2 days

Shipping Expedited across US-Canada:
FedEx, UPS, USPS
Shipping world-wide available:
DHS, FedEx International
Request Quick Quote HVD3102A-NOACC
Equipment info

The Teledyne LeCroy HVD3102A-NOACC is a high-voltage differential probe engineered for precise measurement of high-voltage signals in power electronics and circuit design applications. Operating at 25 MHz bandwidth with a 1 kV CAT safety rating, this probe delivers safe, accurate floating measurements across demanding, noise-rich environments. The dual attenuation architecture—100:1 and 1000:1 ratios—enables flexible measurement ranges from 27.6 V to 1500 V differential, with maximum measurable voltage of 1750 V typical before saturation. Its exceptional common-mode rejection ratio suppresses noise across a broad frequency spectrum, critical for switch-mode power supply characterization and electrical apparatus testing up to 5 kV class.

Technical Specifications

• Bandwidth: 25 MHz
• Rise Time: 14 ns
• Voltage Rating: 1 kV CAT
• Differential Voltage Range (High Attenuation, 100:1): 1500 V (DC + peak AC) from 7 to 500 V/div with up to 1500 V offset
• Differential Voltage Range (Low Attenuation, 1000:1): 27.6 V (DC + peak AC) from 100 mV/div to 6.9 V/div with up to 150 V offset
• Maximum Measurable Differential Voltage: 1750 V (typical)
• Gain Accuracy: 1%
• Input Impedance: 10 MΩ
• Input Capacitance: 2.5 pF
• Attenuation Ratios: 100:1, 1000:1
• Coupling: AC and DC

– Key Features

• Exceptional CMRR across broad frequency range for high common-mode noise rejection
• AutoZero functionality with auto disconnect switch calibrates offset drift without probe removal
• High offset capability at both attenuation settings enables measurement flexibility
• ProBus connector provides integrated power and communication with automatic scaling
• 2 m cable with 0.4 m lead length

– Typical Applications

• 120/240 V switch-mode power supplies
• 600 V class electrical apparatus
• 5 kV class electrical apparatus
• High-voltage floating measurements in noisy power electronic environments

– Compatibility & Integration

• ProBus interface with automatic scaling eliminates external power supplies or batteries
• Designed for Teledyne LeCroy oscilloscope instruments
• Wide oscilloscope compatibility across the LeCroy platform

MPN

HVD3102A-NOACC

Frequency Range

DC to 25 MHz

Voltage / Current Range

27.6 V to 1500 V differential (depending on attenuation); maximum 1750 V typical

Brand Name

Teledyne LeCroy