The HP/Agilent 6114A is a precision DC voltage source engineered for calibration, standards work, and high-performance laboratory applications. It delivers 0–20 V at 2 A or 20–40 V at 1 A through automatic dual-range operation, eliminating manual switching. Voltage control is achieved via pushbutton settings and fine-adjust knob with 200 µV resolution, while front-panel current control enables constant-current operation with 2 mA resolution.
Technical Specifications
DC Output
• Range 1: 0–20 V, 0–2 A
• Range 2: 20–40 V, 0–1 A
• Output Voltage Accuracy: 0.025% + 1.0 mV (at 23 ± 3°C, after 5-minute warm-up)
• Voltage Resolution: 200 µV
• Load Regulation (CV): 0.0005% + 100 µV
• Source Regulation (CV): 0.0005% + 100 µV
• Temperature Coefficient (CV): 0.0001% + 15 µV/°C
• Drift (CV): 0.0015% + 15 µV per 8 hours; 0.0075% + 30 µV per 90 days
Constant Current
• Current Control Resolution: 2 mA
• Load Regulation: 0.01% + 500 µA
• Source Regulation: 0.005% + 40 µA
• Temperature Coefficient: 0.02% + 50 µA/°C
Ripple & Noise (20 Hz–20 MHz)
• Constant Voltage: 100 µV rms / 200 µV peak-to-peak
• Constant Current: 200 µA rms / 1 mA peak-to-peak
Transient Response
• Load Transient Recovery: Less than 50 µs to within 50 mV after full-range current change
– Key Features
• Automatic dual-range operation—no manual switching required
• SCR crowbar overvoltage protection with adjustable limits (0.5 V to 10% above rated output)
• Output impedance: typically 0.05 mΩ series with 3 mH
– Remote Programming
Analog programming via external voltage or resistance. Up-programming speed: 1.75 seconds (full load). Down-programming speed: 350 milliseconds (no load). Compatible with HP Multiprogrammer and HP-IB Isolated D/A Power Supply Programmer.